RP-HPLC Method for Rhein Quantification in Cassia fistula L. (Fabaceae) Leaves

期刊名:Acta Medica Philippina
文献编号:
文献地址: https://actamedicaphilippina.upm.edu.ph/index.php/acta/article/view/8803
发表日期:2024-02-12
Objectives. The aim of this study is to establish a Reversed Phase – High Performance Liquid Chromatographic
(RP-HPLC) method for the quantification of Rhein from Cassia fistula L. leaves.

Methods. A Shimadzu system equipped with a C18 Column (150 x 4.6 mm, 5 μm) with an isocratic elution of
Acetonitrile (solvent A) and 0.1% trifluoroacetic acid aqueous solution (solvent B) (Merck, 1.08178.0050) with a
55:45 ratio, respectively and a flow rate of 1.0 mL/min and sample injection of 10 μL detection was done at 230 nm. Standard solution of Rhein (Chengdu Biopurify) was prepared for method development. This study was validated using the guidelines set under “ICH Topic Q2 R2 or the Validation of Analytical Procedures”. Procedures for linearity, precision, accuracy, limit of detection, and limit of quantitation were performed.

Results. The retention time of Rhein standard was determined at 5.10 minutes. LOD and LOQ were determined to be 1.278 mcg/mL and 3.872 mcg/mL, respectively with good linearity (R2 ≥0.996) with a linear range of 2.5-20 ug/mL of the Rhein standard. The accuracy of the method was determined based on % recovery method and ranged from 94.75%-100.32% (intraday, n=3) with %RSD of 0.71. The intraday precision %RSD was 2.92 (n=6) while interday precision %RSD was 3.75 (n=3). The method was able to check the Rhein quantity among 10 samples of Cassia fistula L. leaves from different locations in the Philippines.

Conclusion. The method was found to be sensitive and accurate for the quantification of Rhein. The method was found to be useful for the quantification of the amount of Rhein and can be used as a Quality Control tool for the assessment of Cassia fistula.